4.9.4
Calibration data
You can examine the calibration factors determined during the calibration routine (SCcal)
and the value currently being used (SCwkg) from this display. If the two values are
different, this indicates a change in instrument temperature between the time of
calibration and the current temperature. Press one of the arrow keys to examine the
temperature information.
Temperature has a small effect on the filter/gas cell characteristics compensated for by the
instrument. Temperature measurement is made within the receiver.
4.9.5
Fault condition
To display the current fault condition, press ENTER while this is displayed.
The instrument automatically selects this display mode if a fault condition occurs. The
instrument recognized the following fault conditions:
1.
ALL CLEAR: No fault condition.
2.
Det. Saturated: The detector level gain within either the receiver or the control unit
is too high for the current duct conditions.
3.
Low Det. Level: Detector levels are too low (< 3,000).
4.
Mod. Freq.O.R.: Chopper motor frequency is out of range (< 30 Hz or > 45 Hz).
5.
Reference Fail.: No reference signal from the source unit.
6.
Cal. Fact. O.R.: After the calibration routine, the calculated Set Cal factor is out of
range. Refer to
Press the arrow key to observe the previous fault condition.
Note
If a fault condition exists, the instrument does not update the minutes, hours, and days averages.
Refer to
Operation
60
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...