Source Intensity Trim Pot
Figure 3-6:
A. Trim pot
3.6.2
Chopper frequency
Complete the following steps to adjust the analyzer's chopper frequency.
1.
Unscrew and remove the source from the air purge.
2.
Unscrew and pull the window plate from the front of the source unit.
. A trim pot (VR1) allows you to adjust the frequency of the chopper
motor.
Configuration and startup
34
Rosemount CCO 5500