Gain Adjust Potentiometers
Figure 3-5:
A. Trim pot
B. Test point
C. LED
D. D2 trim pot
E. D1 trim pot
F. Plant
G. Analog
H. Receiver
I. Serial data
2.
Enter Mode 5
→
Set Detectors and display the values of D2 and D1. Turn the trim pot
counterclockwise until the D2 level is between 12,000 and 15,000. Allow time
between adjustments for the readings to settle.
Note
If the duct is operating and the opacity levels are high, reduce the D2 level to about 8,500.
This should prevent saturation should the opacity level drop off.
Configuration and startup
32
Rosemount CCO 5500