11.
Secure the source unit to the air purge using the four screws removed in
.
Tighten the screws evenly.
12.
Power up and recalibrate the Rosemount CCO 5500 Analyzer.
Refer to
.
5.2.4
Replace receiver unit gas cell
Replace the receiver unit gas cell according to the following procedure.
1.
Shut down and lock out power to the Rosemount
™
CCO 5500 Analyzer.
2.
Unbolt and remove the receiver unit from its air purge.
3.
Loosen the receiver unit cable gland so the cable can slip when the window plate is
removed.
4.
Unscrew and pull the window plate (
) from the front of the receiver unit.
Receiver Gas Cell
Figure 5-8:
A. End detector PCB
B. Window plate
C. M6 bolts
D. M4 x 16 screws
E. Side detector PCB
F. Gas cell assembly
5.
Use a marker pen to match-mark all components of the receiver stack.
6.
Remove the four M4x16 screws that secure the side detector PCB to the receiver
stack.
Maintenance
Reference Manual
87