2.
Span Adjust
3.
Calibrate
Set detectors
You can display both the D1 and D2 levels; you can also display saturation counts. To give
an immediate response to any alterations that are required, the filters and gas cells are not
moved during this operation.
for a discussion of the detector level and saturation count.
Span adjust
Note
Rosemount
™
initially sets the span factor at the factory; do not adjust it unless the instrument
sensitivity is suspected. In any case, Rosemount recommends that you record the orginal value
before making adjustments.
You can adjust instrument sensitivity if a known concentration of gas exists between the
source and receiver units and instrument sensitivity is supsected. If a problem arises,
consult Rosemount.
You may need to adjust the span factor if you have fitted new gas cells or filters.
Calibrate
Re-enter the Mode 5 CALIBRATE menu and proceed to the
Calibrate
option.
You can calculate the basic calibration of the instrument from this routine. It is preferable
to conduct this operation with the plant shut down to ensure a zero gas concentration
within the duct. If this is not possible, the instrument can calibrate to a known value of the
gas concentration - the calibration target.
Configuration and startup
Reference Manual
47
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...