Power Supply
Figure 3-1:
A. AC power selector switch
B. Rail indication LED
4.
Power up the Rosemount CCO 5500 Analyzer and verify that the power supply rail
indication LED (
5.
Install and fasten the power supply unit cover.
6.
Check that the LCD display is functioning at the control unit.
While the source unit is warming up, the LCD display shows
WAITING FOR REFERENCE
.
When the source unit reaches an adequate temperature for the reference to be detected,
the LCD display shows the message
STABILIZING REF
, along with the frequency and
for further details.
The reference frequency takes some time to stabilize (about five minutes from cold
startup). When the reference frequency is within tolerance for 10 consecutive
measurement cycles, the instrument automatically changes to the
OPERATING
mode. This
is Mode
1
and is indicated by a number
1
appearing in the top left corner of the LCD
display. The display shows a reading in ppm; this is not an accurate reading until all
configuration and startup procedures are completed.
Before conducting the alignment procedure, allow 30 minutes for the source temperature
to become stable.
3.4
Alignment
For the instrument to operate properly, the source and receiver units must be aligned.
Rosemount
™
has built in a degree of optical redundancy; normal duct movements do not
affect the operation of the instrument. Read and understand this entire procedure before
starting the alignment.
Configuration and startup
26
Rosemount CCO 5500
Summary of Contents for Rosemount CCO 5500
Page 4: ......
Page 8: ...Contents iv Rosemount CCO 5500 ...
Page 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Page 32: ...Install 24 Rosemount CCO 5500 ...
Page 88: ...Operation 80 Rosemount CCO 5500 ...
Page 98: ...Maintenance 90 Rosemount CCO 5500 ...
Page 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Page 110: ...Spare parts 102 Rosemount CCO 5500 ...
Page 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Page 123: ...Return equipment to the factory Reference Manual 115 ...