6.
Repeat
for the side detector D1 measuring across the test points.
7.
When the detector levels are satisfactory, replace the cover.
Note
If the duct is operating and a high opacity may be in the path, reduce the set voltages to 2 V
rms maximum. This should prevent saturation should the opacity level drop off.
3.5.2
Control unit gain adjustment
After you have optimized the detector level(s) at the receiver, optimize the levels within
the microprocessor. Make the adjustment with two trim potentiometers in the control
unit.
1.
Set the gain to a minimum by turning the D2 detector trim pot (
clockwise.
D2 is a 20-turn potentiometer.
Configuration and startup
Reference Manual
31