Advanced Meter Verification Messages
(continued)
Table 12-3:
Error message
Potential cause
Corrective action
Check for parallel paths in the current loop
Transmitter failure
Perform transmitter self test
Perform manual analog loop test and D/A trim if
necessary
Replace the electronics board
Continuous Meter Verifi-
cation Error
Transmitter calibration verification
test failed
Verify pass/fail criteria
Run manual SMART Meter Verification (8714i) un-
der no flow conditions
Verify calibration using 8714D Calibration Standard
Perform digital trim
Replace electronics stack
Sensor calibration test failed
Run manual SMART Meter Verification (8714i)
Perform sensor tests - see
Sensor coil circuit test failed
Run manual SMART Meter Verification (8714i)
Perform sensor tests - see
Sensor electrode circuit test failed
Run manual SMART Meter Verification (8714i)
Perform sensor tests - see
Verify electrode resistance has a signature value
from a full pipe baseline
Simulated Velocity Out
of Spec
Unstable flow rate during the verifica-
tion test or noisy process
Run manual transmitter verification test with no
flow and a full pipe
Transmitter drift or faulty electronics
Verify transmitter electronics with 8714D Calibra-
tion Standard. The dial on the 8714D should be set
to 30 ft/s (9.14 m/s). The transmitter should be set
up with the nominal calibration number
(1000015010000000) and 5 Hz coil drive frequen-
cy.
Perform an electronics trim using the 8714
If the electronics trim doesn't correct the issue, re-
place the electronics
Coil Resistance Out of
Spec
Moisture in the terminal block of the
sensor or shorted coil
Perform sensor tests - see
If the problem persists, replace the sensor
Coil Signature Out of
Spec
Moisture in the terminal block of the
sensor or shorted coil
Perform sensor tests - see
If the problem persists, replace the sensor
Calibration shift caused by heat cy-
cling or vibration
Perform sensor tests - see
If the problem persists, replace the sensor
Electrode Resistance Out
of Spec
Moisture in the terminal block of the
sensor
Perform sensor tests - see
If the problem persists, replace the sensor
Troubleshooting
Reference manual
161
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...