Set the test criteria for the empty pipe condition. The factory default for this value is set to
five percent with limits configurable between one and ten percent. This parameter applies
to manually initiated test only.
Continuous limit
LOI menu path
Diagnostics > Advanced Diag > Meter Verif > Test Criteria > Con-
tinual
Set the test criteria for the continuous SMART Meter Verification diagnostic. The factory
default for this value is set to five percent with limits configurable between two and ten
percent. If the tolerance band is set too tightly, under empty pipe conditions or noisy
flowing conditions, a false failure of the transmitter test may occur.
9.10
Run manual SMART Meter Verification
LOI menu path
Diagnostics > Advanced Diag > Meter Verif > Run Meter Ver
The SMART Meter Verification diagnostic will be available if the advanced diagnostic suite
(DA2) was ordered. If DA2 was not ordered or licensed, this diagnostic will not be available.
This method will initiate the manual meter verification test.
9.10.1
Test conditions
LOI menu path
Diagnostics > Advanced Diag > Meter Verif > Run Meter Ver >
Test Condition
SMART Meter Verification can be initiated under three possible test conditions. This
parameter is set at the time that the sensor baseline or SMART Meter Verification test is
manually initiated.
No flow
Run the SMART Meter Verification test with a full pipe and no flow in the line. Running the
SMART Meter Verification test under this condition provides the most accurate results and
the best indication of magnetic flowmeter health.
Flowing full
Run the SMART Meter Verification test with a full pipe and flow in the line. Running the
SMART Meter Verification test under this condition provides the ability to verify the
magnetic flowmeter health without shutting down the process flow in applications when a
shutdown is not possible. Running the diagnostic under flowing conditions can cause a
false test failure if there is significant process noise present.
Advanced Diagnostics Configuration
Reference manual
123
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...