9.11.1
Test scope
Continuous SMART Meter Verification can be configured to monitor the sensor coils,
electrodes, analog output, and transmitter calibration, All of these parameters can be
individually enabled or disabled. These parameters apply to continuous SMART Meter
Verification only.
Coils
LOI menu path
Diagnostics > Diag Controls > Cont Meter Ver > Coils
Continuously monitor the sensor coil circuit by enabling this continuous SMART Meter
Verification parameter.
Electrodes
LOI menu path
Diagnostics > Diag Controls > Cont Meter Ver > Electrodes
Continuously monitor the electrode resistance by enabling this continuous SMART Meter
Verification parameter.
Transmitter
LOI menu path
Diagnostics > Diag Controls > Cont Meter Ver > Transmitter
Continuously monitor the transmitter calibration by enabling this continuous SMART
Meter Verification parameter.
Analog output
LOI menu path
Diagnostics > Diag Controls > Cont Meter Ver > Analog Output
Continuously monitor the analog output signal by enabling this continuous SMART Meter
Verification parameter.
9.12
SMART Meter Verification test results
If the SMART Meter Verification test is manually initiated, the transmitter will make several
measurements to verify the transmitter calibration, sensor calibration, coil circuit health,
and electrode circuits health. The results of these tests can be reviewed and recorded on
the
form. Print the "Manual Calibration Verification Results" form and enter the
test results as you view them. The completed form can be used to validate that the meter
is within the required calibration limits to comply with governmental regulatory agencies.
Advanced Diagnostics Configuration
Reference manual
125
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...