3.
Determine the Device ID.
LOI menu path
Detailed Setup > Device Info > Device ID
4.
Obtain a license key from a local Rosemount representative.
5.
Enter license key.
LOI menu path
Diagnostics > Advanced Diag > Licensing > License Key > Li-
cense Key
6.
Enable Diagnostics.
LOI menu path
Diagnostics > Diag Controls
9.3
Tunable empty pipe detection
The tunable empty pipe detection provides a means of minimizing issues and false
readings when the pipe is empty. This is most important in batching applications where
the pipe may run empty with some regularity. If the pipe is empty, this diagnostic will
activate, set the flow rate to 0, and deliver an alert.
Turning empty pipe on/off
LOI menu path
Diagnostics > Diag Controls > Empty Pipe
The tunable empty pipe detection diagnostic can be turned on or off as required by the
application. The empty pipe diagnostic is shipped turned “On” by default.
9.3.1
Tunable empty pipe parameters
The tunable empty pipe diagnostic has one read-only parameter, and two parameters that
can be custom configured to optimize the diagnostic performance.
Empty pipe (EP) value
LOI menu path
Diagnostics > Variables > Empty Pipe
This parameter shows the current empty pipe value. This is a read-only value. This number
is a unit-less number and is calculated based on multiple installation and process variables
such as sensor type, line size, process fluid properties, and wiring. If the empty pipe value
exceeds the empty pipe trigger level for a specified number of updates, then the empty
pipe diagnostic alert will activate.
Advanced Diagnostics Configuration
Reference manual
113
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...