Use this method to reset the maximum electrode coating value.
9.8
4-20 mA loop verification
The 4-20 mA loop verification diagnostic provides a means of verifying the analog output
loop is functioning properly. This is a manually initiated diagnostic test. This diagnostic
checks the integrity of the analog loop and provides a health status of the circuit. If the
verification does not pass, this will be highlighted in the results given at the end of the
check.
The 4-20 mA loop verification diagnostic is useful for testing the analog output when
errors are suspected. The diagnostic tests the analog loop at five different mA output
levels:
•
4 mA
•
12 mA
•
20 mA
•
Low alarm level
•
High alarm level
9.8.1
Initiating 4-20 mA loop verification
LOI menu path
Diagnostics > Advanced Diag > 4-20mA Verify > 4-20mA Verify
The 4-20 mA loop verification diagnostic can be initiated as required by the application. If
the advanced diagnostics suite 2 (DA2 Option) was ordered, then the 4-20 mA loop
verification diagnostic will be available. If DA2 was not ordered or licensed, this diagnostic
is not available.
9.8.2
4-20 mA loop verification parameters
The 4-20 mA loop verification diagnostic has five read-only parameters plus an overall test
result. It does not have any configurable parameters.
4-20 mA loop verification test result
LOI menu path
Diagnostics > Advanced Diag > 4-20mA Verify > View Results
Shows the results of the 4-20 mA loop verification test as either passed or failed.
4 mA measurement
LOI menu path
N/A
Advanced Diagnostics Configuration
Reference manual
119
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...