8.6.5
Special flow rate unit
LOI menu path
Basic Setup > Flow Units > Special Units > Rate Unit
Flow rate unit is a format variable that provides a record of the units to which you are
converting. The Handheld Communicator will display a special units designator as the
units format for your primary variable. The actual special units setting you define will not
appear. Four characters are available to store the new units designation. The LOI will
display the four character designation as configured.
To display flow in acre-feet per day, and acre-foot is equal to 43560 cubic feet, the
procedure would be:
1.
Set the volume unit to ACFT.
2.
Set the base volume unit to ft3.
3.
Set the conversion factor to 43560.
4.
Set the time base unit to Day.
5.
Set the flow rate unit to AF/D.
Advanced Configuration Functionality
110
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...