Shows the measured value of the 4 mA loop verification test.
12 mA measurement
LOI menu path
N/A
Shows the measured value of the 12 mA loop verification test.
20 mA measurement
LOI menu path
N/A
Shows the measured value of the 20 mA loop verification test.
Low alarm measurement
LOI menu path
N/A
Shows the measured value of the low alarm verification test.
High alarm measurement
LOI menu path
N/A
Shows the measured value of the high alarm verification test.
9.9
SMART
™
Meter Verification
The SMART Meter Verification diagnostic provides a means of verifying the flowmeter is
within calibration without removing the sensor from the process. This diagnostic test
provides a review of the transmitter and sensor's critical parameters as a means to
document verification of calibration. The results of this diagnostic provide the deviation
amount from expected values and a pass/fail summary against user-defined criteria for the
application and conditions. The SMART Meter Verification diagnostic can be configured to
run continuously in the background during normal operation, or it can be manually
initiated as required by the application.
9.9.1
Sensor baseline (signature) parameters
The SMART Meter Verification diagnostic functions by taking a baseline sensor signature
and then comparing measurements taken during the verification test to these baseline
results.
Advanced Diagnostics Configuration
120
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...