Conductivity limits
Process liquid must have a minimum conductivity of 5 microSiemens/cm (5
micromhos/cm) or greater.
Process temperature limits
PTFE lining
–58 to 350 °F (–50 to 177 °C)
ETFE lining
–58 to 300 °F (–50 to 149 °C)
PFA and PFA+ lining
-58 to 350 °F (–50 to 177 °C)
Polyurethane lining
0 to 140 °F (–18 to 60 °C)
Neoprene lining
0 to 176 °F (–18 to 80 °C)
Linatex lining
0 to 158 °F (–18 to 70 °C)
Adiprene lining
0 to 200 °F (–18 to 93 °C)
Temperature vs. Pressure Limits for ASME B16.5 class flanges
Table A-9:
Sensor temperature vs. pressure limits for ASME B16.5 class flanges (
½
Flange material
Flange rating
Pressure
@ -20 to 100 °F
(-29 to 38 °C)
@ 200 °F (93 °C)
@ 300 °F (149 °C)
@ 350 °F (177 °C)
Carbon Steel
Class 150
285 psi
260 psi
230 psi
215 psi
Class 300
740 psi
675 psi
655 psi
645 psi
Class 600
1000 psi
800 psi
700 psi
650 psi
Class 600
1480 psi
1350 psi
1315 psi
1292 psi
Class 900
2220 psi
2025 psi
1970 psi
1935 psi
Class 1500
3705 psi
3375 psi
3280 psi
3225 psi
Class 2500
6170 psi
5625 psi
5470 psi
5375 psi
304 Stainless
Steel
Class 150
275 psi
235 psi
205 psi
190 psi
Class 300
720 psi
600 psi
530 psi
500 psi
1000 psi
800 psi
700 psi
650 psi
1440 psi
1200 psi
1055 psi
997 psi
Class 900
2160 psi
1800 psi
1585 psi
1497 psi
Class 1500
3600 psi
3000 psi
2640 psi
2495 psi
Class 2500
6000 psi
5000 psi
4400 psi
4160 psi
(1) Liner temperature limits must also be considered.
(2) 30-in. and 36-in. AWWA C207 Class D rated to 150 psi at atmospheric temperature.
(3) Option Code C6.
(4) Option Code C7.
(5) Option Code S6.
(6) Option Code S7.
Product Specifications
196
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...