Flow units (PV)
The flow units variable specifies the format in which the flow rate will be displayed. Units
should be selected to meet your particular metering needs. See
.
Line size
The line size (sensor size) must be set to match the actual sensor connected to the
transmitter. The size must be specified in inches.
Upper range value (URV)
The URV sets the 20 mA point for the analog output. This value is typically set to full-scale
flow. The units that appear will be the same as those selected under the flow units
parameter. The URV may be set between –39.3 ft/s to 39.3 ft/s (–12 m/s to 12m/s). There
must be at least 1 ft/s (0.3 m/s) span between the URV and LRV.
Lower range value (LRV)
The LRV sets the 4 mA point for the analog output. This value is typically set to zero flow.
The units that appear will be the same as those selected under the flow units parameter.
The LRV may be set between –39.3 ft/s to 39.3 ft/s (–12 m/s to 12m/s). There must be at
least 1 ft/s (0.3 m/s) span between the URV and LRV.
Calibration number
The sensor calibration number is a 16-digit number generated at the factory during flow
calibration, is unique to each sensor, and is located on the sensor tag.
5.3
Local operator interface (LOI)
To activate the optional LOI, press the DOWN arrow. Use the UP, DOWN, LEFT(E), and
RIGHT arrows to navigate the menu structure. A complete map of the LOI menu structure
is shown in
.
The display can be locked to prevent unintentional configuration changes. The display lock
can be activated through a HART communication device, or by holding the UP arrow for
three seconds and then following the on-screen instructions.
When the display lock is activated, a lock symbol will appear in the lower right hand corner
of the display. To deactivate the display lock, hold the UP arrow for three seconds and
follow the on-screen instructions. Once deactivated, the lock symbol will no longer appear
in the lower right hand corner of the display.
5.4
Field Communicator interface
Use the menu paths to configure basic setup of the transmitter using a field
communicator.
Basic Configuration
46
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...