Optional housing
316/316L unpainted, option code SH
Type 4X and IEC 60529 IP66
Cover gasket
Aluminum housing: Buna-N
316 SST housing: Silicone
Electrical connections
Conduit entries
Available in 1/2 inch NPT or M20. See ordering table footnotes for details
Terminal block screws
6-32 (No. 6) suitable for up to 14 AWG wire
Safety grounding
screws
External stainless assembly, M5; internal 8-32 (No. 8)
Vibration rating
Integral mount
2G per IEC 61298
Remote mount
5G per IEC 61298
Dimensions
See Product Data Sheet.
Weight
Rosemount 8732EM
Aluminum
Approximately 7 lbs. (3.2 kg)
316 stainless steel
Approximately 23 lbs. (10.5 kg)
Add 1 pound (0.5 kg) for local operator interface.
A.3
8705-M Flanged Sensor Specifications
Product Specifications
194
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Summary of Contents for Rosemount 8732EM
Page 2: ......
Page 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Page 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Page 253: ...Implementing a Universal Transmitter Reference manual 247 ...