Table 3-4: Split shroud removal tool list
Tools required
Standard 7/16 inch (11 mm) socket wrench or a standard 7/16 inch wrench
Appropriate personal protective equipment
Figure 3-12: Split shroud removal
A. Upper split shroud
B. Hex nuts with integral tooth lock washers and shroud clamps
C. Lower split shroud
Procedure
1. Using a 7/16 inch (11 mm) wrench, loosen and remove the four lower split shroud
hex nuts with integral tooth lock washers and shroud clamps (see
).
2. Remove the shroud clamps from each of the four studs by pulling the clamp off of
the ‘stud at each location.
3. Slide the lower split shroud down to remove it from the meter body.
4. Move the shroud assembly and bolts out of your immediate work area.
This completes the split shroud removal procedure.
3.4.4
Remove latched band shrouds
The latched band shroud is utilized on 2 inch to 10 inch ultrasonic flow meters.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
50
Emerson.com/Rosemount