11. If replacing other electronics or the fuse, continue with
Replace the I.S. Barrier board
Replace the Power Supply Board
or
Replace the Acquisition Module for direct or remote mount electronics
before
replacing the end caps and sealing the enclosure.
12. If you encounter problems replacing the electronics, see the Emerson Customer
Support contact information on the back cover of this manual.
3.9.2
Replace the fuse
Procedure
1. Disable electrical power to the meter.
2. Refer to
for the tools required to complete this
procedure.
3. Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latch (requires a 3 mm Allen wrench) on the terminal end of the enclosure
(see
) and remove the end cap.
4. Use a 1/4 inch (7mm) standard flat head screw driver and remove the Fuse holder
).
Figure 3-58: Fuse holder cap
A. Fuse holder cap
5. Remove the fuse from the holder.
6. Insert the replacement fuse (Littlefuse #218002.HXP) into the Fuse holder.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
94
Emerson.com/Rosemount