2. Use retaining ring pliers to pull the transducer feed-thru cable seal from the
transducer housing.
Figure 3-15: Remove transducer from housing
A. Feed thru cable seal removal
B. Retaining ring pliers
3. Slide the feed-thru cable seal down the cable.
Figure 3-16: Feed-thru cable seal removal
A. Feed-thru cable seal
4. Use the Retaining Ring pliers and remove the retaining ring, shoulder washer, and
spring from the transducer housing.
5. Pull the transducer capsule from the housing.
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
53