This completes re-wetting the transducer procedure.
3.6
Transducer housings or housing seals
replacement
If it is necessary to remove the transducer housings or housing seals, use the following
steps to safely remove and reinstall them. This procedure applies to meters with direct
mount or remote mount Transmitter Electronics Enclosures.
WARNING
CUTTING HAZARD
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter.
Failure to comply may cause serious injury.
WARNING
FLUID CONTENTS MAY BE UNDER PRESSURE
When the meter is under pressure, DO NOT attempt to remove or adjust the transducer
housing.
Attempting to do so may release pressurized fluid, resulting in serious injury for personnel
or equipment damage.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812