Figure 3-25: Transmitter Electronics Enclosure and Base Enclosure removal
A. Transmitter electronics enclosure
B. Base enclosure bolts
C. Base electronics enclosure
5. Lift the Transmitter Electronics Enclosure off of the Base Enclosure to expose the
Acquisition Module wiring. Prop the Transmitter Electronics Enclosure on top of the
Base Enclosure and prepare to disconnect the Acquisition cable and transducer
terminal blocks.
6. Use a 1/8 inch (3mm) flat blade screw driver to remove the Acquisition cable
terminal block screws. Unplug the Acquisition cable terminal block.
7. Use a 1/8 inch (3mm) flat blade screw driver and remove the transducer wire
terminal blocks screws. Disconnect the transducer cable wires from the terminal
blocks so that the cables can be removed from the meter.
NOTICE
Ensure the transducer cables are labeled for the chord configuration (A1, A2, B1 and
B2).
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
65