Figure 3-32: Transducer removal
A. Feed-thru cable seal
B. Shoulder washer
C. Spring
D. Transducer capsule
Figure 3-33: LT-10 Transducer and cable assembly removed from meter
17. Prepare to install a new transducer pair and the cable assembly.
18. Place one drop of Dow Corning 200 Silicon Oil
1-360-01-650) on the face of the transducer to be installed.
19. Insert the transducer capsule into the transducer housing in the meter body.
20. Slide the spring and shoulder washer along the cable to the transducer capsule.
(5)
Dow Corning 200 is a trademark of Dow Corning Corporation, U.S.A.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
70