NOTICE
Feed-thru foam seals tear easily. Carefully pull the transducer cables through the
seal to prevent ripping the feed-thru cable seal.
27. Reattach the Base Enclosure to the upper split shroud and the meter body with the
three shoulder washers, flat washers, split lock washers and hex head bolts. Tighten
the bolts with a 7/16 inch (11 mm) socket driver.
Figure 3-38: Base Enclosure assembly
A. 7/8 hex head bolts (3)
B. Split lock washers (3)
C. Flat washers (3)
D. Nylon shoulder washers (3)
E. O-ring
F. Base enclosure
28. Apply Loctite
®
nickel anti-seize or equivalent to the three hex head bolts to prevent
loosening due to vibration during operation.
29. Place the Acquisition Module into the Base Enclosure, align the thee holes of the
Acquisition Module with the three holes in the base enclosure and secure with the
three split lock washers and flat head screws (see
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
74