Table 2-2: Maintenance
(continued)
MeterLink utility
Diagnostics
Action(s)
MeterLink Tools Menu
Analog outputs
Run Analog Outputs test and verify outputs are
within 4mA -20mA range
•
0% = 4mA
•
25% = 8mA
•
50% = 12mA
•
75% = 16mA
•
100% = 20 mA
MeterLink Tools Menu
Digital outputs
•
Run Digital Outputs test.
•
Digital Output Content is in relation to
frequency validity and flow direction
configuration and polarity.
Meter Electronics
Acquisition Module
communications error
•
If the CPU Module LED 5 is not flashing
green, check interconnect cable between
Acquisition Module and the CPU Module.
•
Check firmware revision and upgrade if
necessary using
MeterLink Tools
→
Program Download
.
•
If the CPU Board LED 5 is not flashing green,
replace Acquisition Module (see
the Acquisition Module for direct or remote
mount electronics
).
Troubleshooting
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812