2170100001L29 2109V005
5
Important information
NOTICE
i
Do not use damaged image plates Damaged
image plates may not provide reliable diag-
nostic images
i
Laser product compliance: VistaScan Ultra
View meets the requirements of IEC 60825-1
Ed 3 (2014)
i
EMC Compliance Requirements:
i
The unit is intended for use in professional
healthcare facilities (in accordance with IEC
60601-1-2) If the appliance is operated in
another environment, potential effects on
electromagnetic compatibility must be taken
into account
i
Do not operate the unit in the vicinity of HF
surgical instruments or MRT equipment
i
Maintain a minimum distance of at least 30
cm between the unit and other electronic
devices
i
Keep a minimum distance of 30 cm between
the unit and mobile radio devices
i
Stacking or using the scanner adjacent to
other equipment may violate EMC compliance
and interfere with the scanner operation
i
The device transmits and receives RF at
1356 MHz
i
Operate the scanner outside of the patient
environment, at least 15 m away from the
patient Do not touch the device and the
patient simultaneously
i
Image plate preparation: Completely clean
and erase image plates before taking an X-ray
exposure See the image plate preparation
section of this manual
i
Use care when handling image plates:
Avoid fingerprints and scratching Refer
to the instructions provided with the im-
age plate package for further information
on handling
i
Equipment disposal Disposal of VistaScan
Ultra View, including internal electronic circuitry
and image plates must be carried out only at
the appropriate facilities for recovery and re-
cycling Make sure that such items are dis-
posed of in accordance with current federal,
national, state and local government rules
and regulations
i
Keep device dry Do not spray solvents or
liquid directly on the scanner
i
Insert image plate only Avoid damage to the
VistaScan Ultra View transport by making
sure that the light protection cover does not
get pulled into the transport along with the
image plate
i
The equipment EMISSIONS characteristics
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A) If it is used in a
residential environment (for which CISPR
class B is normally required) this equipment
might not offer adequate protection to radio
frequency communication services The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment
1.2 Copyright information
All circuits, processes, names, software
programs and units mentioned in this
document are protected by copyright
The Installation and Operating Instructions
must not be copied or reprinted, neither in
full nor in part, without written authorization
from Dürr Dental
EN
i
Portable RF communications equipment (in-
cluding peripherals such as antenna cables
and external antennas) should be used no
closer than 30 cm to any part of the VistaScan
Ultra View, including cables specified by the
manufacturer Otherwise, degradation of the
performance of this equipment could result