Troubleshooting
EN
Problem
Probable cause
Solution
Error code
10000
Unit is exposed to too much
light
i
Darken the room
i
Turn the unit so that no light is directly
incident into the entry slot
Error code
10009
Internal communication error
warning; unit remains ready for
operation
i
Update the firmware
Error code
10017
Unit shuts down
i
Wait until the unit has shut down
completely
Error code 2
System error during startup of
the unit
i
Switch the unit off and back on again
i
Update the firmware
Error code 78
Memory card full
i
Transmit image data to the computer
i
Insert an empty memory card
Fault during memory cleanup
i
Press and hold the reset button while
switching on the unit
i
Update firmware
i
Press and hold the reset button while
switching on the unit
Firmware not
running
Firmware update has been
carried out
i
Switch the unit off and on again
Internal communication fault
i
Switch the unit off and on again
Settings (eg
language) reset
after unit re-
start
Faulty configuration file
i
Update the firmware
i
Reset the configuration to the factory
settings and reconfigure
Warning
message
during shut-
down of the
unit
Not a malfunction
i
Update the firmware