20
2170100001L29 2109V005
Assembly
EN
Operating system
Microsoft Windows 7 Professional,
Enterprise, or Ultimate with Service Pack 1
for an Intel 32-bit or an Intel 64-bit extended
(x64) processor
Microsoft Windows 81 Professional or
Enterprise for an Intel 64-bit extended (x64)
processor
Microsoft Windows 10 Professional or
Enterprise for an Intel 64-bit extended (x64)
processor
NOTICE
Damage due to user error.
Only qualified specialists or employees
trained by Dürr Dental are permitted to
install, connect and start using the unit
NOTICE
Damage due to physical shock.
During operation and the switching off
process, ensure that the unit is not moved
in any way whatsoever, as the moving parts
must be allowed to stop moving completely
6 Requirements
6.1 Installation/setup room
– VistaScan Ultra View should be set up in a
closed, dry and well-ventilated room
– Room temperature should not fall below
+10°C (+50°F) or 35°C (+95°F)
CAUTION
Danger of short circuit due to
condensation.
VistaScan Ultra View can only be taken
into operation once the unit has warmed
up to room temperature and is dry
– Do not install in a utility room, eg in a boiler
room or in a wet room
– Set up in ordinary room light Avoid direct
sunlight and bright interior lighting (more than
1000 lux) directly above the plate guide
– There must be no source of interference in the
room for set-up (eg strong magnetic field)
Any such disturbances can interfere with the
operation of the VistaScan Ultra View
In order to avoid problems when scan-
ning the image data, VistaScan Ultra
View should be set up so that it will not
be subject to knocks, vibration or similar
6.2 Counter top/table set-up
– Provide a stable, flat counter top or table large
enough for the scanner work area and able to
support the weight of the unit (see 4 Techni-
cal data)
– Locate VistaScan Ultra View no further than
1,8 m from a socket outlet
– Make sure to leave the socket outlet easily
accessible, so that the unit may be unplugged
if necessary The power supply is the mains
disconnect device
– Portable and mobile HF communication ap-
pliances can interfere with the effectiveness
of electrical medical devices
– Do not arrange the unit so that it is im-
mediately next to, or stacked with other
appliances However, if the unit is operated
immediately next to, or stacked with other
appliances, make sure to check the unit
carefully in the configuration selected to
verify normal operation
6.3 Monitor
The monitor must comply with the requirements
for digital X-ray with a high light intensity and
wide contrast range
Strong ambient light, sunlight falling directly onto
the monitor and reflections can make it harder
or even impossible to perform a diagnosis
based on the X-ray images
6.4 Computer system requirements
The system requirements for the com-
puter systems can be found in the down-
load area at wwwduerrdentalcom (doc-
ument no 9000-618-148)