2170100001L29 2109V005
51
EN
15 Scanning times
The scanning time corresponds to the time taken for complete scanning of image data and depends
on image plate format and pixel size
The time to image will depend largely on the computer system used and its work load Times stated
are approximate
Theoretical
resolution (LP/mm)
40
25
20
10
Pixel size (μm)
12.5
20
25
50
Intra Size 0 (2 x 3)
26 s
16 s
13 s
6 s
Intra Size 1 (2 x 4)
32 s
20 s
16 s
8 s
Intra Size 2 (3 x 4)
32 s
20 s
16 s
8 s
Intra Size 3 (27 x 54)
40 s
25 s
20 s
10 s
Intra Size 4 (57 x 76)
53 s
33 s
27 s
14 s
16 File size (uncompressed)
The actual file size will depend on the image plate format and the pixel size File sizes stated are ap-
proximate and have been rounded upwards
Suitable compression methods can considerably reduce the file size without loss of data
Theoretical
resolution (LP/mm)
40
25
20
10
Pixel size (μm)
12.5
20
25
50
Intra Size 0 (2 x 3)
986 MB
385 MB
246 MB
062 MB
Intra Size 1 (2 x 4)
1229 MB
480 MB
307 MB
077 MB
Intra Size 2 (3 x 4)
1627 MB
636 MB
407 MB
102 MB
Intra Size 3 (27 x 54)
1901 MB
743 MB
475 MB
119 MB
Intra Size 4 (57 x 76)
5545 MB
2166 MB
1386 MB
347 MB
Appendix