Important information
i
Note that cable lengths and cable extensions
have effects on electromagnetic compatibility
i
No maintenance measures are required to
maintain the EMC basic safety
NOTICE
Negative effects on the EMC due to
non-authorised accessories
i
Use only Dürr Dental parts or acces-
sories specifically approved by Dürr
Dental
i
Using any other accessories may result
in increased electromagnetic interfer-
ence emissions or the unit having re-
duced electromagnetic immunity, lead-
ing to an erroneous operation mode
NOTICE
Erroneous operation mode due to use
immediately adjacent to other devices
or with other stacked devices
i
Do not stack the unit together with
other devices
i
If this is unavoidable, the unit and oth-
er devices should be monitored in or-
der to ensure that they are working
correctly
NOTICE
Reduced performance character-
istics due to insufficient distance
between unit and portable HF com-
munication devices
i
Keep a distance of at least 30 cm be-
tween the unit (including parts and
cables of the unit) and portable HF
communication devices (wireless
units) (including their accessories
such as antenna cables and external
antennas)
2.7 Essential performance charac-
teristics
The VistaScan Ultra View unit does not possess
any significant performance characteristics as set
out in IEC 60601-1 (EN 60601-1) section 43
The unit complies with the requirements according
to IEC 60601-1-2:2014
EN
2.8 Notification requirement for se-
rious incidents
The user / patient is required to report to the
manufacturer and the competent authority of the
Member State in which the user and/or patient is
established any serious incident that has oc-
curred in relation to the device
2.9 Only use original parts
i
Only use Dürr Dental parts or accessories and
optional articles specifically approved by Dürr
Dental Only use original wear parts and re-
placement parts
Dürr Dental accepts no liability for dam-
ages or injury resulting from the use of
non-approved accessories or optional
accessories, or from the use of non-origi-
nal wear parts or replacement parts
The use of non-approved accessories,
optional articles or non-genuine wear
parts / replacement parts (eg mains ca-
ble) can have a negative effect in terms of
electrical safety and EMC
2.10 Transport
The original packaging provides optimum pro-
tection for the unit during transport
If required, original packaging for the unit can be
ordered from Dürr Dental
Dürr Dental will not accept any responsi-
bility or liability for damage occurring dur-
ing transport due to the use of incorrect
packaging, even where the unit is still un-
der guarantee
i
Only transport the unit in its original packaging
i
Keep the packing materials out of the reach of
children
i
Do not expose the unit to any strong vibra-
tions or shocks
2.11 Disposal
An overview of the waste keys for Dürr Dental
products can be found in the download area at
wwwduerrdentalcom (document no
P007100155)
Unit
The unit must be disposed of properly
Within the European Union, the unit must
be disposed of in accordance with EU
Directive 2012/19/EU (WEEE)
If you have any questions about the correct
disposal of parts, please contact your dental
trade supplier