40
2170100001L29 2109V005
Usage
EN
12 Cleaning and disinfection
When cleaning and disinfecting the unit and its
accessories, observe country-specific directives,
standards and specifications for medical prod-
ucts as well as the specific specifications for
dental practices and clinics
NOTICE
The use of unsuitable agents and
methods can damage the unit and
accessories.
Do not use any products based on phe-
nolic compounds, halogen-releasing
compounds, strong organic acids or ox-
ygen-releasing compounds, as they
may damage the materials
i
Dürr Dental recommends using disin-
fectants from the Dürr Dental product
range Only the products specified in
these instructions have been subject-
ed to material compatibility testing by
Dürr Dental
i
Read the operating instructions for
the disinfectants
Wear protective gloves
12.1 Image plate scanner
Unit surfaces
The unit surfaces must be cleaned and disin-
fected of any contamination or visible soiling
Dürr Dental recommends using the disinfectants
FD 322, FD 333, FD 350 and FD 366 sensitive
NOTICE
Liquid can cause damage to the unit.
i
Do not spray the unit with cleaning
and disinfectant agents
i
Make sure that liquid does not get in-
side the unit
i
Remove any soiling with a soft, lint-free cloth
that has been dampened with cold tap water
i
Disinfect the surfaces using a disinfection
wipe Alternatively, use disinfectant on a soft,
lint-free cloth
12.2 Light protection cover
The surface of the unit must be cleaned and
disinfected if it is contaminated or visibly soiled
i
Disinfect the light protection cover using a dis-
infectant before and after placement Dürr
Dental recommends FD 333 forte wipes (viru-
cidal), FD 350 (limited virucidal activity) and FD
322 premium wipes (limited virucidal activity)
i
Allow the light protection cover to completely
dry before using it
12.3 Image plate
Cleaning and disinfection wipes are unsuitable
for cleaning image plates and may cause dam-
age to them
Only use a cleaning agent that is compatible
with the materials:
Dürr Dental recommends the image plate clean-
ing wipe (see "34 Consumables") Only this
product has been subjected to material compat-
ibility testing by Dürr Dental
NOTICE
Heat or humidity will damage the im-
age plate.
i
Do not steam sterilise the image plate
i
Do not immersion-disinfect the image
plate
i
Only use cleaning agents that are
compatible with the materials
12.4 Plate guides
The plate guides must be cleaned and disinfect-
ed if there are indications of contamination or
visible dirt
Durr Dental recommends using the disinfectants
FD 322, FD 333, FD 350, and FD 366 sensitive
Use the following cleaning and disinfecting im-
mersion disinfectants:
i
ID 213 Instrument disinfection
i
ID 212
i
ID 212 forte
NOTICE
Heat can damage plastic parts.
i
Do not use a thermal disinfector or
steam sterilizer on any device