34
2170100001L29 2109V005
Usage
EN
11 Operation
11.1 Changing the plate guide.
i
VistaScan Ultra View can be used to scan im-
age plates Size 0, Size 1, Size 2, Size 3 and
Size 4
i
Each size of image plate requires the match-
ing size plate guide
i
The size of the image plate is clearly marked
on the plate guide
i
Remove the plate guide by simply lifting it
from the transport
i
Install the plate guide by aligning the position
pins and placing the guide into the guide
holes It will snap into place due to the magnet
installed
CAUTION
If a plate guide is removed, a warning
message will appear on the LCD screen
that the plate guide is missing
11.2 X-ray
The procedure is described using a size
2 VistaScan Image Plate IQ S2 as an ex-
ample
Required accessories:
– VistaScan Image Plate IQ
– VistaScan Light Protection Cover Plus of the
same size as the image plate
– Plate guide and image plate are same size
WARNING
Risk of cross contamination when
not using the light protection cover or
when using the light protection cover
more than once
i
Do not use an image plate without a
light protection cover
i
Do not use the light protection cover
more than once (disposable item)
Preparing the X-ray
– The image plate has been cleaned
– The image plate is not damaged
– The RFID tag is attached to the image plate
If the RFID tag peels off, replace the image
plate
i
If using it for the first time or if it has been
stored for over a week: erase the image plate
(see "115 Erasing the image plate")