2170100001L29 2109V005
55
19 Recommended separation distances between equipment
Distance between portable and mobile RF communications equipment
VistaScan Ultra View is intended for use in an electromagnetic environment in which radiated RF inter-
ferences are controlled The user of the VistaScan Ultra View unit can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and VistaScan Ultra View as recommended below, in accordance with the
maximum output power of the communications equipment
Maximum assigned
output power of the
transmitter
W
Separation distance as a function of the transmitter's frequency
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 25 GHz
d = 116 √P
d = 116 √P
d = 233 √P
001
0116
0116
0233
01
0366
0366
0736
1
116
116
233
10
366
366
736
100
116
116
233
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in metres (m) can be established by using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to
the transmitter manufacturer
Note 1:
At 80 and 800 MHz, the higher frequency range applies
Note 2:
These guidelines may not apply in all situations
Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people