17 Electromagnetic Compatibility (EMC) Compliance Information
The following provides the EMC technical data used to show that each VistaScan Ultra View is designed
and manufactured to meet the electromagnetic requirements of international standard IEC 60601-1-2
as shown below The user should ensure that the VistaScan Ultra View unit is indeed used within this
environment to ensure that the system will perform as intended
i
Portable and mobile RF communications equipment can affect VistaScan Ultra View operation
i
Use of accessories or cables other than those specified or provided by the manufacturer may result
in increased VistaScan Ultra View emissions and decreased VistaScan Ultra View immunity
i
The VistaScan Ultra View should not be used adjacent to or stacked with other equipment that may
interfere with proper VistaScan Ultra View operation
18 Manufacturer’s Guidance and Declaration - Electromagnetic
Emission
s
The VistaScan Ultra View is intended to be used within the electromagnetic environment specified be-
low The user of the VistaScan Ultra View should ensure that it is used in such an environment
Emissions test
Compliance
Electromagnetic
Environment - Guidance
RF emissions
ClSPR 11
Group 1
VistaScan Ultra View uses radio-electric energy only
for its internal subsystems It therefore emits very
low energy and is not likely to interfere with nearby
electronic devices
RF emissions
ClSPR 11
Class A
VistaScan Ultra View is suitable for use in all establish-
ments other than domestic, and may also be used in
domestic establishments and those directly connected
to the public low voltage power supply network that sup-
plies buildings used for domestic purposes provided the
following warning is heeded:
Warning:
This equipment is intended for use by health-
care professionals only This equipment may cause radio
interference or may disrupt the operation of nearly equip-
ment It may be necessary to take mitigation measures,
such as reorienting or relocating the VistaScan Ultra View
or shielding the location
Harmonic emissions
EN 61000-3-2
Class A
Voltage fluctuations/flicker
EN 61000-3-3
Complies