Troubleshooting
Problem
Probable cause
Solution
Ghosting or dou-
ble exposure
on X-ray image
Image plate exposed twice
i
Only expose the image plate once
Image plate not properly erased
i
Check the erasure unit for proper func-
tion
i
Inform a service technician if the problem
persists
X-ray image mir-
rored in one corner
Image plate bent during X-ray ex-
posure
i
Do not bend the image plate
Shadow on the X-
ray image
Image plate removed from the light
protection cover before scanning
i
Do not handle image plates without a light
protection cover
i
Store the image plate in a light protection
cover
X-ray image cut
off, part missing
A metal part of the X-ray tube is in
front of the X-ray beam
i
Recording an X-ray image, make sure
there are no metal parts between the X-
ray tube and the patient
i
Check X-ray tube
Faulty edge masking in imaging
software
i
Deactivate edge masking
Software unable
to combine the
data to make a
complete image
The X-ray dose on the image plate
was insufficient
i
Increase X-ray dose
Unsuitable scanning mode se-
lected
i
Select a suitable scanning mode
Amplification (HV value) is set too
low in the software
i
Increase amplification (HV value)
The setting for the threshold value
is too high
i
Reduce the threshold value