Troubleshooting
Problem
Probable cause
Solution
X-ray image has
strips on image
Image plate has been pre-exposed,
eg by natural radiation or stray X-
ray radiation
i
If the image plate has been stored for
more than one week, erase the image
plate prior to use
Parts of image plate exposed to
light during handling
i
Do not expose exposed image plates to
bright light
i
Scan image data within half an hour after
the exposure
Image plate dirty or scratched
i
Clean the image plate
i
Replace scratched image plate
Bright stripes in the
scanning window
Too much incident ambient light
during the scanning process
i
Darken the room
i
Turn the unit so that no light is directly in-
cident on the input unit
Horizontal, grey
lines in the X-ray
image, extending
beyond the left
and right image
edge
Transport slipping
i
Clean the transport mechanism, replace
the transport belts if necessary
X-ray image is
stretched length-
wise with bright,
horizontal stripes
Wrong light protection cover or im-
age plate used
i
Only use original accessories