26
2170100001L29 2109V005
Assembly
EN
8.5 X-ray unit settings
Intraoral X-ray units
If 60 kV can be set on the X-ray unit, this setting is preferred
The standard exposure values for F-speed film (eg Kodak Insight) can be used
The following table shows the standard values for the exposure time and the dose area product of an
image plate for an adult patient The information dealing with the duration of X-ray exposition and the
dose area product is only applicable when VistaIntra is used as X-ray device
DC emitter 60kV/7mA Tube Length 20cm
Without X-rayfield limitation
X-ray field limitation 2x3 X-ray field limitation 3x4
60kV
mGycm²
µGy
60 kV
mGycm²
µGy
60 kV
mGycm²
µGy
Incisors
010 s
183
5387 010 s
38
5464 010 s
77
5499
Premolars
014 s
256
7668
01 s
54
7662 014 s
108
7689
Molars
019 s
348
1037
019 s
73
1040
019 s
147
1040
Bite wing
020 s
366
1090
020 s
77
1100
020 s
155
1095
i
Check and adjust the specific X-ray unit in accordance with the standard values