Supplement – Infinity
®
Delta Series Software VF10.1
67
Supplement for Infinity Delta Series Software
During a measurement, if p-IT or p-BT readings become invalid or go out of range, the averaging
screen disappears. Check that p-ITand p-BT are in range before restarting the measurement.
[p-IT range: -3 to 31 °C (27 to 88 °F),
p-BT
range: 25 to 43 °C (77 to 109 °F)]
Peripheral Devices and Associated Software
Analysis Tool
When receiving data from the Perseus A500, or Atlan A350/300, if the Analysis Tool is open and the
A500,A350, or A300 exits standby, the VTi parameter label is not displayed. Close and reopen the
Analysis Tool to restore the parameter label.
Overview
This chapter describes interfaces that can be used to connect the bedside monitor to peripheral devices
and supported ventilators and anesthesia systems. This chapter also describes the Analysis tool and the
Independent Surgical Display.
A monitor with compatible software displays waveforms and parameter values acquired from peripheral
devices via the general Medical Information Bus (MIB) interface protocol or the MEDIBUS.X interface
protocol. These protocols significantly expand the monitor's capabilities. The MIB protocol and the
MEDIBUS.X protocol require the monitor to be connected to an Infinity Docking Station (IDS).The
MEDIBUS.X interface protocol allows the peripheral device to be connected directly to the IDS. The MIB
interface protocol requires the MIB II protocol converter to connect the device to the IDS.
The Delta series monitor displays only a subset of the parameters supported by these devices but relays
all supported parameters to the infinity network and networked devices.
CAUTION
Recalibration of the thermodilution measurement is recommended at least every eight hours and with
significant changes in hemodynamic conditions such as volume shifts or changes to medication.
NOTE
“Analysis tool” replaces “Open Lung tool” throughout these instructions for use.
NOTE
MIB and MEDIBUS.X functionality is not available when the monitor is in DirectNet mode.
NOTE
Refer to the RS-232 export handbook for information on which parameters are available for export
protocol. The RS-232 export handbook is available in English only.