Supplement – Infinity
®
Delta Series Software VF10.1
91
Supplement for Infinity Delta Series Software
EMC declaration
General information
This device was tested for electromagnetic compatibility using accessories from the list of accessories.
Other accessories may only be used if they do not compromise the electromagnetic compatibility. The use
of non-compliant accessories may result in increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of other devices only if Dräger has approved this device
arrangement. If no approval has been given by Dräger, it must be ensured that this device functions
correctly in the desired arrangement before use. The instructions for use for the other devices must be
followed.
Electromagnetic environment
NOTE:
• For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Emissions
Compliance
Radiated emissions
Kappa, Infinity Docking Station (IDS), integrated
MIB, SCIO Four Module, BISx pod:
– Class A, group 1 (30 MHz to 1 GHz)
Delta, Delta XL:
– Class B, group 1 (30 MHz to 1 GHz)
Conducted emissions
Kappa, Infinity Docking Station (IDS), integrated
MIB, SCIO Four Module, BISx pod:
– Class A, group 1 (150 kHz to 30 MHz)
Delta, Delta XL:
– Class B, group 1 (150 kHz to 30 MHz)