Supplement for Infinity Delta Series Software
24
Supplement – Infinity
®
Delta Series Software VF10.1
Arrhythmia Monitoring
Arrhythmia Ranges and Defaults
Parameter
Alarm Priority
Default
Rate (default)
Count (default)
Alarm Archive Factory
Default Setting
ASY
HIGH
Not adjustable
Not adjustable
Str./Rec.
VF
HIGH
Not adjustable
Not adjustable
Str./Rec.
VT
HIGH
100 to 200 (120
)
Increments of 10
5 to 15 (
1
0)
Increments of 1
Str./Rec.
RUN
MED
Not adjustable (VT
rate)
Not adjustable (3 to
VT rate - 1)
Str./Rec.
AIVR
MED
Not adjustable = (VT
rate - 1)
Not adjustable (3)
OFF
SVT
MED
120 to 200 (150
)
Increments of 10
3 to 10 (3)
Increments of 1
Str./Rec.
TACH
MED
100 to 200 (130
)
Increments of 10
5 to 15 (8)
Increments of 1
OFF
BRDY
MED
30 to 105 (50)
Increment of 5
Not adjustable (8)
OFF
CPT
LOW
Not adjustable
Not adjustable
Str./Rec.
BGM
LOW
Not adjustable
Not adjustable
Str./Rec.
PAUS
OFF
1 to 3.5 (2.5)
Increments of 0.5
Not adjustable
OFF
ARTF
OFF
Not adjustable
Not adjustable
OFF
NOTE
When in neonatal mode or pediatric mode with Basic Arrhythmia monitoring, bradycardia is a high priority
heart rate alarm that is adjustable.