Supplement – Infinity
®
Delta Series Software VF10.1
99
Supplement for Infinity Delta Series Software
Continuous Non-Invasive Arterial Blood Pressure via CNAP Pod
Static cuff accuracy
±3 mmHg (± 0.4 kPa)
Calibration range
Adult and Pediatric:10 to 260 mmHg ±3 mmHg (1.3 to 34.7 kPa ±
0.4 kPa)
Neonatal: 10 to 150 mmHg ±3 mmHg (1.3 to 20.0 kPa ± 0.4 kPa)
Parameter label
CNAP
Systolic
CNAP-S
Diastolic
CNAP-D
Mean
CNAP-M
NOTE:
Only values for systolic, diastolic, and
mean are displayed in the p-box. Labels are not
displayed.
Units
mmHg (kPa)
Measuring range
Systolic
40 to 250 mmHg (5.3 to 33.3 kPa)
Diastolic
30 to 210 mmHg (4 to 28 kPa)
Mean
35 to 230 mmHg (4.66 to 30.7 kPa)
HR range (for CNAP measurements)
20 to 200 bpm
Accuracy
± 5 mmHg (0.6 kPa)
Display resolution
1 mmHg (0.1 kPa)
Inflation pressure
Default
120 mmHg (16 kPa)
Minimum
30 mmHg (4 kPa)
Maximum
300 mmHg (40 kPa)
Over pressure limit
Pressure
300 mmHg (40 kPa) + 10%
Response time
≤
3 s
Deflation time
≤
15 s (to reach < 15 mmHg)