Supplement for Infinity Delta Series Software
36
Supplement – Infinity
®
Delta Series Software VF10.1
WARNING
If the SpO
2
sensor is used in the presence of intravascular dyes, such as methylene blue,
measurement accuracy may be reduced.
Do not rely on measurement data if the SpO
2
sensor is used under these conditions.
WARNING
Risk of inaccurate data
In the presence of shock, low blood pressure, severe vasoconstriction, major anemia,
hypothermia, arterial occlusion proximal to the sensor, and asystole, the pulse rate may be
inaccurate.
WARNING
Risk of patient injury
High intrathoracic pressure, pressure on the thorax and other consecutive impairments of the
venous flow can lead to venous pulsation and pulse signal failure.
Do not position the SpO
2
sensor where it might be affected in this way.
WARNING
To reduce the hazard of burns during surgery, keep the sensor or transducer and their
associated cables away from the surgical site, the electro-surgical unit return cathode, and the
earth ground.
CAUTION
For Masimo SET pod, after connection of the sensor, observe monitor for any messages if the sensor does
not light up. If the sensor LED does not light up, replace sensor and/or intermediate cable.
CAUTION
Refer to the “Pulse Oximetry (SpO
2
)” chapter in the
Infinity Delta Series Instructions for Use for Software
VF9
for additional precautions regarding safe device operation.
CAUTION
Do not immerse the sensor or patient cable in any liquid. Moisture may present a safety risk.
NOTE
When in neonatal mode or pediatric mode with Basic Arrhythmia monitoring, bradycardia is a high heart
rate alarm that is adjustable.