Supplement for Infinity Delta Series Software
92
Supplement – Infinity
®
Delta Series Software VF10.1
Recommended separation distances from wireless communication devices
To ensure that the functional integrity of this device is maintained, there must be a separation distance of
at least 1.0 m (3.3 ft) between this device and wireless communication devices.
NOTE
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required), this equipment might not offer adequate protection to radio-frequency communication ser-
vices. The user might need to take mitigation measures, such as relocating or re-orienting the equip-
ment.
Immunity against
Test level and required electromagnetic environ-
ment
Electrostatic discharge (ESD) (IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances (bursts)
(IEC 61000-4-4)
Power cable: ±2 kV
Longer signal input lines/output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5)
Voltage, external conductor – external conductor:
±1 kV
Voltage, external conductor – protective ground con-
ductor: ±2 kV
Magnetic fields at mains frequency
(IEC 61000-4-8)
30 A/m
Voltage dips and short interruptions in the supply
voltage (IEC 61000-4-11)
100 % drop, 0.5 period
100 % dip, 1 period
30 % dip, 25/30 periods
Voltage interruptions:
100 % drop, 5 seconds
Radiated high-frequency disturbances
(IEC 61000-4-3)
80 MHz to 2.7 GHz: 3 V/m
Conducted high-frequency disturbances
(IEC 61000-4-6)
150 kHz to 80 MHz: 3 Vrms, ISM bands: 6 Vrms
Electromagnetic fields in the vicinity of wireless
communication devices
Various frequencies from 385 MHz to 5785 MHz:
9 V/m to 28 V/m, with various pulse modulations