Supplement – Infinity
®
Delta Series Software VF10.1
75
Supplement for Infinity Delta Series Software
Atlan, Primus, Zeus, Apollo, and Perseus Anesthesia Devices
The Delta/Delta XL/Kappa monitor displays data from the devices listed at the beginning of this chapter.
When the monitor is connected to an Infinity Primus, Apollo, Perseus,or Atlan anesthesia machine, it
displays the concentrations of CO
2,
O
2,
N2O, and anesthetic agents (Halothane, Isoflurane, Enflurane,
Sevoflurane, or Desflurane). The "Anesthesia Parameters" table lists all displayed parameters.
The monitor uses the multigas parameter box, etCO2* parameter box, and waveform channels to display
parameter values, waveforms, and loops. The agent parameter box displays agent concentrations for
anesthetic gases.
Parameter values acquired from the anesthesia machine are trended in graphical and tabular form.The
anesthesia machine provides the range and resolution for all parameters.
Parameter settings, alarms, and error messages are not displayed on the monitor. They are displayed on
the network device such as an Infinity CentralStation, Innovian, Gateway, or Symphony instead. See the
Instructions for Use for your network device for the list of parameters transferred from the anesthesia
machine. Refer to the Primus, Apollo, Perseus,and Atlan instructions for use for details on calculating
parameter values.
NOTE
The Infinity Primus and the Zeus anesthesia machines are not sold in the United States.
NOTE
The bedside monitor displays the xMAC (MAC multiple) values only if available and sent by the
connected anesthesia device.