Supplement for Infinity Delta Series Software
90
Supplement – Infinity
®
Delta Series Software VF10.1
Recommended separation distances
Immunity test
IEC 60601 test level:
Compliance
level:
Recommended separation
distance:
Conducted RF
rf coupled into lines
(IEC 61000-4-6)
150 kHz to 80 MHz
3 Vrms
Radiated rf
(IEC 61000-4-3)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3 V/m
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in metres (m).
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the equipment is used exceeds the applicable RF
compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the equipment.
2
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
d
1.2
∗
P
=
d
1.2
∗
P
=
d
2.3
∗
P
=
Recommended separation distances between portable and mobile RF communications
equipment and the equipment
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter in meters
150 kHz - 80 MHz
80 MHz to 800MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
d
1.2
∗
P
=
d
1.2
∗
P
=
d
2.3
∗
P
=