Supplement for Infinity Delta Series Software
116
Supplement – Infinity
®
Delta Series Software VF10.1
Pulse Oximetry (SpO
2
)
Quick Reference Table – SpO
2
Setup
If double-pulsing occurs when
Pulse Tone Volume
is enabled under the SpO2
parameter menu, disable
the pulse tone volume for SpO2 and enable the pulse tone volume in the
ECG Options
menu.
Dräger Mainstream etCO2 Monitoring
Precautions
Multiple CO2* Devices
In a multiple etCO2* devices situation involving MIB and Scio, it is recommended to physically disconnect
the undesired device from the monitor.
When the message
Duplicate Device Connected
appears and if the CO2* waveform label disagrees with
the parameters being monitored (e.g., the “CO2* module” waveform label displays when MIB is the active
CO2* source device), disconnect one of the CO2* devices, wait a few second, then reconnect the CO2*
source device.
NOTE
If a MultiGas parameter box appears after a Restore Setup with a Dräger mainstream CO2 module
connected to the monitor, do the following:
1
Disconnect the Dräger mainstream CO2 module.
2
Connect a Scio module.
3
Disconnect the Scio module.
4
Reconnect the Dräger Mainstream CO2 module.
This procedure should remove the MultiGas parameter box.