Status Messages
Message
Possible Cause
Suggested Action
NBP Cannot
Measure
The pulse profile is too poor to make a
reliable measurement (usually due to
persistent motion artifact).
• Check the patient and treat, if
necessary.
• Move the cuff to a limb with less
movement.
• Restart measurement. If the
message persists, contact the
hospital’s technical personnel or
DrägerService.
• Take a manual measurement if the
condition persists.
NBP
Measurement
Timeout
A measurement was aborted (usually
due to motion artifact) because it lasted
longer than two minutes (adult or
pediatric) or 90 seconds (neonatal).
• Repeat the measurement.
• Take a manual measurement if the
condition persists.