Supplement for Infinity Delta Series Software
60
Supplement – Infinity
®
Delta Series Software VF10.1
Neuromuscular Transmission (NMT) Monitoring
Precautions
Patient Preparation
Careful skin preparation is important to ensure appropriate electric contact for stimulation. Follow the
clinical techniques approved at your hospital for preparing the patient's skin prior to administering the
electrodes.
MultiGas Reduced
Accuracy
Warm-up phase in progress.
NOTES
– Measured parameter labels
display but are followed by a
“?”.
– Alarm history displays
MultiGas Reduced
Accuracy
. When this
condition is removed, the
alarm history displays
MultiGas Reduced
Accuracy Ended
.
– Parameter box (Pbox)
question marks (?) are not
transmitted/displayed over
the network. They only
appear in Pboxes.
– Wait until the warm-up phase has ended
before relying on the displayed MultiGas
values for clinical evaluation.
– If
Auto Zero Delay
has been selected,
allow the auto zero to occur.
– If the problem persists, call
DrägerService.
MultiGas Zero
Accepted
The zeroing cycle was
successful.
No action is required.
Message
Condition
Suggested action
WARNING
To avoid serious injury to the patient, apply the stimulation electrodes close together as
described. Do not apply the electrodes
transcerebrally (across the head), directly over the eyes, covering the mouth, on the front of the
neck (particularly over the carotid sinus), trans-thoracically (across the chest), on the upper
back, or other position(s) that could place the electrodes over the heart. Doing so can cause
electrical current to enter the chest or head and may lead to irregularities in cardiac rhythm or
brain activity, or to pain.