Supplement for Infinity Delta Series Software
84
Supplement – Infinity
®
Delta Series Software VF10.1
To clean the Infinity PiCCO Pod
1
Disconnect the Infinity PiCCO pod from the bedside monitor.
2
Clean the Infinity PiCCO pod with a gauze pad moistened with soapy water or with an approved
cleaning agent.
3
Dry thoroughly with a lint-free cloth.
To disinfect the Infinity PiCCO Pod
1
Disconnect the Infinity PiCCO pod from the bedside monitor.
2
Disinfect the Infinity PiCCO pod with a gauze pad moistened with diluted alcohol.
3
Dry thoroughly with a lint-free cloth.
etCO2
Dräger Mainstream CO
2
Module, Sensor, and Cuvettes
Mainstream Module
To clean the Mainstream Module
1
Disconnect the sensor cable (SpO
2
cable, if using) from the Dräger etCO2 mainstream module.
2
Wipe the module surfaces with a cloth moistened with a soap solution.
3
Dry thoroughly with a lint-free cloth.
To disinfect the Mainstream Module
1
Disconnect the sensor cable (including the SpO
2
cable, if using) from the Dräger etCO2 Mainstream
Module.
2
Wipe the module surfaces with a gauze moistened with one of the following:
– A 1:3 solution of alcohol
– A 1:10 solution of sodium hypochlorite (household bleach)
3
Dry thoroughly with a lint-free cloth.
Mainstream Sensor and Cuvettes
For cleaning and disinfecting instructions for the sensor and reusable cuvettes, refer to the Dräger Infinity
MCable – Mainstream CO
2
IFU.
WARNING
To reduce the risk of infection, remember that the disposable cuvettes are for single patient use
only and cannot be sterilized.