Supplement for Infinity Delta Series Software
34
Supplement – Infinity
®
Delta Series Software VF10.1
– Perform device checks (for example, calibration of pod with cuff controller, sensor cuff, and NBP
cuff)
– Check/Visual inspection
3
Check that the device combination is in good condition:
– All labels are complete and legible
– There is no visible damage
4
Use the instructions for use to check that all components and accessories needed to use the product
are available.
5
Check the electrical safety requirements according to IEC 62353.
Metrological checks
If required by applicable regulations, the following measurement functions must be checked every two
years by qualified DrägerService personnel:
– Body temperature
– Non-invasive blood pressure
Preventive Maintenance
The following table shows the preventive maintenance intervals:
WARNING
Risk of faulty components
Device failure is possible due to wear or material fatigue of the components. To maintain proper
operation of all components, this device must undergo inspection and preventive maintenance
at specified intervals.
WARNING
Risk of electric shock
Before performing any maintenance work, disconnect all electrical connectors from the power
supply.
Component
Interval
Measure
Responsible Personnel
CNAP SmartPod
Every 2 years
Check
Expert
CNAP cuff controller
Every 2 years
Check
Expert
Air pump
Every 2 years
Replace
Expert